DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
8 - 90 Section 8-7 - Replacing OPIO
8-7-3 Replacement of the Probe Holder Set-T
This table includes information specific to these instructions. For more information 8-2-6 "Tools needed
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6
8-7-3-1 Removal Procedure
8-7-3-2 Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8-
12 on page 8-223
Table 8-39 Manpower, Time, Tools, and PPE
Manpower Time Tools PPE
One person Approximately 5 minutes No tools required
Cut resistant gloves
Knee Pad
Step Corresponding graphic
1.
Press the button underneath the T2 Side Probe Holder Left and
slide to the left to remove from the console
2.
Press the button underneath the T2 Side Probe Holder Right and
slide to the right to remove from the console.