GE HealthCare CONFIDENTIAL
DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
8 - 98 Section 8-7 - Replacing OPIO
8-7-10 Replacement of the UOPIO-T Assy
This table includes information specific to these instructions. For more information 8-2-6 "Tools needed
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6
8-7-10-1 Preparations
1.) Remove the OPIO Top Cover Set-T as described in 8-7-2 "Replacement of the OPIO Top Cover
Set-T" on page 8-89
Table 8-46 Manpower, Time, Tools, and PPE
Manpower Time Tools PPE
One person Approximately 10 minutes Philips Screwdriver
Cut resistant gloves
Knee Pad