EasyManuals Logo

GE LOGIQ Totus User Manual

GE LOGIQ Totus
644 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #272 background imageLoading...
Page #272 background image
GE HealthCare CONFIDENTIAL
Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
7 - 22 Section 7-3 - Service Desktop
7-3-6 DICOM
Use DICOM to troubleshoot DICOM components and connectivity. Specifically, use these pages:
• DICOM Service Status
• DICOM Spooler
For more information :
• "DICOM Service Status" on page 7-22
• "DICOM Spooler" on page 7-24
DICOM Service Status
DICOM Service Status displays information on the DICOM components and connectivity. Use DICOM
to test the connectivity of all the available DICOM devices or individual DICOM devices. In addition, you
can:
• View DICOM details
• View DICOM connectivity status
• Stop a service
• Ping a single service
• Ping all services
DICOM connectivity status displays as green or red.
• Green means that the DICOM ping passed
• Red means that the DICOM ping failed
The information on DICOM Service Status is available to Class C and Class M licenses.
To access DICOM Service Status, navigate to Utility (second page) > Service > DICOM > DICOM
Service Status.
Figure 7-12 DICOM Service Status

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ Totus and is the answer not in the manual?

GE LOGIQ Totus Specifications

General IconGeneral
BrandGE
ModelLOGIQ Totus
CategoryMedical Equipment
LanguageEnglish

Related product manuals