Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
7 - 22 Section 7-3 - Service Desktop
7-3-6 DICOM
Use DICOM to troubleshoot DICOM components and connectivity. Specifically, use these pages:
DICOM Service Status
DICOM Spooler
For more information :
"DICOM Service Status" on page 7-22
"DICOM Spooler" on page 7-24
DICOM Service Status
DICOM Service Status displays information on the DICOM components and connectivity. Use DICOM
to test the connectivity of all the available DICOM devices or individual DICOM devices. In addition, you
can:
View DICOM details
View DICOM connectivity status
Stop a service
Ping a single service
Ping all services
DICOM connectivity status displays as green or red.
Green means that the DICOM ping passed
Red means that the DICOM ping failed
The information on DICOM Service Status is available to Class C and Class M licenses.
To access DICOM Service Status, navigate to Utility (second page) > Service > DICOM > DICOM
Service Status.