5 - 28 Section 5-9 - Top Console Description
5-9-3 Control Panel Movement - Principle of Operation
The Control panel can be rotated, adjusted in height.
• Rotation Adjustment : 30°
• Height Adjustment : 250mm (9.8 inch)
CAUTION
To avoid injury or damage, make sure nothing is within the range of motion before moving the
operator panel. This includes both objects and people.
NOTICE
!! NOTICE:
Do NOT attempt to apply excessive rotating force to the operator panel. The elevation brake is
designed to hold the operator panel in desired position during normal use, and not meant to completely
fix the operator panel position.