GE HealthCare CONFIDENTIAL
DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
Section 8-7 - Replacing OPIO 8 - 113
8-7-20-2 Removal Procedure
8-7-20-3 Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8-
12 on page 8-223
Step Corresponding graphic
1. Unscrew the nut to remove the OPIO Mode Encoder Set-T.
2.
Pull the OPIO Mode Encoder Set-T to remove from the
OPIO PWA-T.