GE HealthCare CONFIDENTIAL
DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
8 - 12 Section 8-4 - Loading / Reloading / Upgrading the Software
8-4-7 Loading the base image and application software - full load (continued)
Read and understand the contents of this procedure before performing the software update.
Before loading the base image and application software, you must perform the following steps:
Back up patient data by using EZBackup or Export.
NOTE: Do not use Patient Archive Backup/Restore in Utility > System > Backup Restore to back up
patient data. For more information on backing up and restoring patient data or system presets,
“Backup and Restore” in the LOGIQ Totus™ User Manual, Direction 5929163-1EN.
Back up user-defined configurations only. Leave service unchecked.
NOTE: Use a USB thumb drive to back up user-defined configurations. These configurations will be
restored once the software load is complete.
For software options (for example, Breast Assistant, Powered by Koios DS (Utility > Connectivity>
Summary > Koios), Tricefy (Utility > Connectivity > Summary > Tricefy), and Device
Management (Utility > Connectivity > Summary > Device Mgmt) capture the information on
those screens before loading the full ghost.
NOTE: Once the software installation is complete, use this information to set up the options again.
Confirm that no images are in the clipboard or opened exams.
Disconnect the Ultrasound system from the network and remove all transducers from the system.
If used, disconnect the secondary monitor.
Connect and turn ON the printer to retain printer preferences. Before you load the base image,
record any printer preferences. After the base image is loaded, check the printer preferences and
update as needed.
If a printer is plugged in at this time, it is automatically installed.
You must record all service presets on paper to restore remote connectivity after software
installation is complete. You will have to manually restore service presets (for example, Agent
Configuration).
From Windows, Utility > Connectivity > Network > Wired, Wireless and Proxy or from the
customer, manually record the DNS server settings.

Table of Contents

Related product manuals