DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
Section 8-4 - Loading / Reloading / Upgrading the Software 8 - 13
8-4-7 Loading the base image and application software - full load (continued)
Table 8-6 Load Base Image and Application Software- full load
Step Corresponding Graphic
1.
Remove any USB flash drives and any CD/DVDs from
the DVD drive prior to updating the software.
Disconnect the system from the network and remove all
transducers from the system.
If used, disconnect the secondary monitor, any HDMI or
DisplayPort cables and/or external monitors.
Power down the LOGIQ Totus.
Insert the software thumb drive into the USB port on the
back of the LOGIQ Totus.
Power on the LOGIQ Totus.
2
CAUTION
Selecting 1- (FULL) is a DATA WIPE & System Application reload. ALL EXISTING PATIENT DATA WILL BE
REMOVED. All existing data will be removed. System and application installation will be completed without
preserving customer data.
With the trackball, point to 1 - (FULL) and then press the
left set key to select
1 - (FULL).
With the trackball, point to OK and press the left set key
to select OK.
In the Warning dialog box, select Yes.
This will take approximately 10-15 minutes.