GE HealthCare CONFIDENTIAL
Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
Chapter 7 Diagnostics/Troubleshooting 7 - 17
7-3-5 Home (continued)
This table shows all the elements available on System Information with descriptions.
For more information :
• "Home" on page 7-15
Table 7-2 System Information
Element Description
CRM Number
Customer Relationship Management (CRM) number. System identifier assigned to
the customer unit by the service region.
Agent Registered
Registered status of the agent. Valid values are:
• Yes – The agent is registered in the back office.
• No – The agent is not registered in the back office.
• Not Available – The agent is not running or has not been configured.
Agent Quarantine
Quarantine status of the agent. Valid values are:
• Yes – The agent has more than one device registered with the same CRM Number
in the back office.
• No – The agent has one device registered with the listed CRM Number in the back
office.
• Not Available – The agent is not running or has not been configured.
Agent CRM Verified
CRM verified status of the agent. Valid values are:
• Yes – The agent is verified in the back office.
• No – The agent is not verified in the back office.
• Not Available – The agent is not running or has not been configured.
Agent Connected
Connection status of the agent. Valid values are:
• Yes – The agent is connected to the back office.
• No – The agent is not connected to the back office.
Not Available – The agent is not configured or running.
Model Number
GE HealthCare part number for the LOGIQ Totus. The same number as listed on the
rating plate.
Serial Number Serial number of the LOGIQ Totus. The same number as listed on the rating plate.
System Type Product name of the LOGIQ Totus.
Facility Name of the hospital or facility where the LOGIQ Totus is installed.