EasyManuals Logo

GE LOGIQ Totus User Manual

GE LOGIQ Totus
644 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #275 background imageLoading...
Page #275 background image
GE HealthCare CONFIDENTIAL
Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
Chapter 7 Diagnostics/Troubleshooting 7 - 25
7-3-6 DICOM (continued)
This table shows all the elements available on DICOM Spooler with descriptions.
For more information :
• "DICOM" on page 7-22
• "DICOM Service Status" on page 7-22
Table 7-7 DICOM Spooler
Element Description
Job Status Displays the number of jobs for each type by status. For example, Pending.
Spooler Health
DICOM Spooler Space allotted to DICOM jobs on the E: drive.
Free Space Available space on the E: drive.
Total Size of the E: Drive Total space on the E: drive.
Spooler Summary
Job ID Number for the job.
Job Status
Current status of the job. These are the valid values:
• Pending
• Active
• Hold
• Done
• Success
• Failed
Job Type
Type of job. These are the valid values:
• Storage – Images
• Structured Reports
• Key Image Notes
• STC – Storage Commitment
• MPPS – Modality Performed Procedure Step
Destination Name of service receiving the objects.
Age Amount if time the job has been in the spooler.
Spooler Entry time Time the job entered the spooler.
Transfer time Time the job was transferred from the spooler.
Job Size Size of the job.
Error details If applicable, details for the error.