Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
Chapter 7 Diagnostics/Troubleshooting 7 - 39
7-3-7 Utilities (continued)
To view the Class M license information:
1.) Insert the SSA key.
2.) Navigate to Utility (second page) > Service > Utilities > SSA License.
3.) Under Class M key Details, view the values. For example, the SSO ID for the user assigned to the
SSA key.
If Not Available displays for all of these values, the SSA key is not validating.
To restore an SSA key that is not validating:
1.) Remove the SSA key from the LOGIQ Totus.
2.) If open, close the Service desktop.
3.) Restart the LOGIQ Totus.
4.) Once the LOGIQ Totus has restarted, plug in the SSA key.
5.) Navigate to Utility (second page) > Service > Utilities > SSA License.
6.) Click Restore SSA.
7.) Check to if the SSA key validates.
For more information :
"Utilities" on page 7-26
"When running Scan Assistant and pausing or editing annotations/comments, the annotation/
comments might remain on screen or may fail to display. To fix this problem, press the Comment
key twice to close or display annotation/comments." on page 7-76