Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
7 - 72 Section 7-8 - Probes Troubleshooting
7-8-5 Automatic probe diagnostics (VITAOnDemand) (continued)
To run a VITAOnDemand probe test:
1.) From Scan, navigate to the second page of the B-mode tab.
2.) Select Probe Test. A popup window displays requesting to clean the probe before starting the test.
3.) Clean the probe.
4.) Select Yes to start the test.
5.) If the test passes, a popup window will show that the probe test passed.
6.) If the test fails, a popup window displays to suggest next steps to retest the probe.