DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
8 - 24 Section 8-4 - Loading / Reloading / Upgrading the Software
8-4-8 Loading the base image and application software - System only (continued)
Loading application software
Read and understand the contents of this procedure before performing the software update.
Before loading the base image and application software, you must perform the following steps:
Back up patient data and system presets.
Confirm that no images are in the clipboard or opened exams.
Disconnect the system from the network and remove all transducers from the system.
If used, disconnect the secondary monitor.
Connect and turn ON the printer to retain printer preferences. Before you load the base image,
record any printer preferences. After the base image is loaded, check the printer preferences and
update as needed.
If a printer is plugged in at this time, it is automatically installed.
Save all service presets to restore remote connectivity after software installation is complete.
Save all user presets.
NOTE: For more information on backing up and restoring patient data or system presets, Backup and
Restore” in the LOGIQ Totus Basic User Manual (latest revision).
WARNINGWARNING
Before loading software, CONFIRM WITH THE CUSTOMER OR SITE ADMINISTRATOR if the
patient archive has been encrypted and if the customer has the password or the key to unlock
the data. Failure to obtain those items, will cause permanent DATA LOSS!
NOTICE
Before loading the base image load software, remove all media from USB ports and internal drives (for
example, DVD), and disconnect any portable hard drives. Failing to do so may interrupt the software
load process.
Table 8-9 Loading the base image and application software - System only
Step Corresponding Graphic
1.
Suspend encryption prior to turn off the system.
Suspended encryption will be maintained until reboot,
make sure this is the last step prior to reload.
Remove any USB flash drives and any
CD/DVDs from the DVD drive prior to updating the
software.
Power down the LOGIQ Totus.
Insert the software thumb drive into the USB port on the
back of the LOGIQ Totus.
Power on the LOGIQ Totus.