GE HealthCare CONFIDENTIAL
DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
8 - 26 Section 8-4 - Loading / Reloading / Upgrading the Software
5.
View these screens as the application software loads.
This will take approximately 15-20 minutes.
BasePatches
• Shutting down running programs
• Updating policy files
• Updating drivers
• Applying Security Patches
Service Platform
LOGIQ Software Installation
• Removing previous installation
• Installing application
• Configuring application
• Updating Files
• Service Platform Prerequisites
6.
Once the application software installation is complete, the
LOGIQ Totus restarts.
7.
After boot up, log in as ADM (no password required) and, when prompted, select Lowest for the security level.
NOTE: Select the Lowest security level and then continue with installation and testing. After the update is complete, you can
transfer the LOGIQ Totus to the Site Administrator who can reset the security level for the site. The security level can be reset
through Utility > Admin > Logon.
8.
Go to "Full disk encryption and replacement procedures" on page 8-37.
Table 8-9 Loading the base image and application software - System only(Continued)
Step Corresponding Graphic