GE HealthCare CONFIDENTIAL
DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
8 - 42 Section 8-5 - Replacing Cables
CAUTION
Use caution to ensure that the power cable does not disconnect during system use.
If the system is accidentally unplugged, data may be lost.
WARNINGWARNING
To avoid leakage current above safety limits as prescribed by IEC 60601-1 and to ensure continuity of protective
earth, DO NOT connect the Ultrasound system and mains-operated accessories to a single or multiple socket
extension cord or power strip.
4.
Unscrew 2 screws to remove the power cable clamp.
5.
Remove the power cable.
Table 8-20 Power cable removal