GE HealthCare CONFIDENTIAL
DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
Section 8-5 - Replacing Cables 8 - 51
8-5-8-2 Removal Procedure
Steps Corresponding Graphic
1. Unlock Cable Clamp.
2. Unscrew 2 screws to remove the Cable Clamp.
3. Cut 2 Cable Ties.
4. Pull UBE To OPIO Touch Panel HDMI Cable to
remove from the T2 OPIO BASE.
5. Unscrew 4 screws to remove the Unity Cable Cover
Brkt