Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
Chapter 10 Care & Maintenance 10 - 21
Section 10-6
Electrical Safety Tests
10-6-1 Safety Test Overview
The electrical safety tests in this section are based on and conform to ANSI/AAMI ES60601-1(For USA)
and IEC60601-1 Medical Equipment Safety Standards. They are intended for the electrical safety
evaluation of cord-connected, electrically operated, patient care equipment. If additional information is
needed, refer to the ANSI/AAMI ES60601-1(For USA) and IEC 60601-1 documents.
To minimize the risk of a probe causing electrical shock, the customer should observe the following
recommendations:(twice a year is recommended)
• DO NOT use a probe that is cracked or damaged in any way
• Check probe leakage current:
* once a year on surface probes (or in accordance with your facilities QA program)
* once a year on endocavitary probes (or in accordance with your facilities QA program)
* whenever probe damage is suspected
WARNINGWARNING
!! WARNING:
THE USER MUST ENSURE THAT THE SAFETY INSPECTIONS ARE PERFORMED AT
LEAST EVERY 12 MONTHS ACCORDING TO THE REQUIREMENTS OF THE PATIENT
SAFETY STANDARD AAMI/ANSI ES 60601-1 AND IEC60601-1. ONLY TRAINED
PERSONS ARE ALLOWED TO PERFORM THE SAFETY INSPECTIONS MENTIONED
ABOVE.
CAUTION
!! CAUTION:
To avoid electrical shock, the unit under test must not be connected to other electrical
equipment. Remove all interconnecting cables and wires. The unit under test must not be
contacted by users or patients while performing these tests.
CAUTION
!! CAUTION:
Possible risk of infection. Do not handle soiled or contaminated probes and other components
that have been in patient contact. Follow appropriate cleaning and disinfecting procedures
before handling the equipment.
WARNINGWARNING
!! WARNING:
Test the system, peripherals and probes for leakage current. Excessive leakage current
can cause FATAL INJURY OR DEATH. High leakage current can also indicate
degradation of insulation and a potential for electrical failure. DO NOT use probes or
equipment having excessive leakage current.