Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
Chapter 10 Care & Maintenance 10 - 35
Type BF Patient Leakage Current Test - Probes (This test should be done for all probe)
Type BF Patient Leakage Current Test - Mains to Probe (This test should be done for all probes)
Final Check. All system covers are in place. System scans with all probes as expected.
Accepted by: ______________________________________________________________________
10-7-6 Ultrasound Equipment Quality Check (EAC and IQC)
Download and use the latest version of these forms. They can be retrieved from MyWorkshop.
• EQC -- Refer to DOC0929340 in MyWorkshop.
Table 10-21
Tester
Neutral
Tester Polarity
Switch
Ground
Switch
Max Value
Allowed
Value
Measured
OK? Comments
Closed Normal Closed 0.1mA
Closed Normal Open 0.5mA
Closed Reversed Closed 0.1mA
Closed Reversed Open 0.5mA
Open Normal Closed 0.5mA
Open Reversed Closed 0.5mA
Table 10-22
Tester
Neutral
Tester Polarity
Switch
Ground
Switch
Max Value
Allowed
Value
Measured
OK? Comments
Closed Normal Closed 5mA
Closed Reversed Closed 5mA