Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
Chapter 2 Site Preparations 2 - 3
2-2-1 LOGIQ Totus environmental requirements (continued)
Cooling
The cooling requirement for the LOGIQ Totus with monitor and onboard peripherals is up to 3070 Btu/
hr. This figure does not include cooling needed for lights, people, or other equipment in the room. Each
person in the room places an additional 300 Btu/hr. demand on the cooling system.
Lighting
Bright light is needed for LOGIQ Totus set up, updates, and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interference.