ADVANCED OPERATION
147
9. Press Return[F7] to return to the previous menu
and to set the sweep options:
Sweep[F5]
Sets the sweep to Single
or Cont. When set to
Single the Sweep
Single[F6] soft-key must
be pressed to initiate a
sweep.
Sweep Single[F6]
Initiates a single sweep
when Sweep[F5] is set to
Single.
10. Press Decoding[F3] to set the line decoding
method. The decoded data can be seen when in
the Symbol or Waveform View:
None[F1]
No line decoding of
data.
Miller[F2]
Miller decoding (aka.
Delayed Encoding).
Manchester
Encoding[F3]
Manchester encoding
(aka. Phase Encoding).