Page #328 background imageLoading...
Page #328 background image
GSP-9300 User Manual
326
GSP-9300 Specifications
The specifications apply when the GSP is powered on for 45
minutes * to warm-up to a temperature of 20˚C to 30˚C, unless
specified otherwise.
* 45 minutes typical, 90 minutes maximum
Frequency
Frequency
Range
9 kHz to 3.0 GHz
Resolution
1 Hz
Frequency Reference
Accuracy
±(period since last adjustment X aging rate) +
stability over temperature + supply voltage
stability
Aging Rate
±2 ppm max.
1 year after last
adjustment
Frequency
Stability over
Temperature
±0.025 ppm
0 to 50 °C
Supply Voltage
Stability
±0.02 ppm
Frequency Readout Accuracy
Start, Stop,
Center, Marker
±(marker frequency indication X frequency
reference accuracy + 10% x RBW + frequency
resolution
1
)
Trace points
Max 601 points, min 6 points
Marker Frequency Counter
Resolution
1 Hz, 10 Hz, 100 Hz, 1 kHz
Accuracy
±(marker frequency
indication X frequency
reference accuracy +
counter resolution)
RBW/Span >=0.02 ;
Mkr level to DNL>30
dB
Frequency Span
Range
0 Hz (zero span), 100
Hz to 3 GHz
Resolution
1 Hz
Accuracy
± frequency resolution
1
RBW: Auto;

Table of Contents

Other manuals for GW Instek GSP-9300

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GSP-9300 and is the answer not in the manual?

GW Instek GSP-9300 Specifications

General IconGeneral
BrandGW Instek
ModelGSP-9300
CategoryMeasuring Instruments
LanguageEnglish