GSP-9300 User Manual
136
Note
The MAX and MIN measurements are held until
higher or lower values are found. To reset the MAX
and MIN measurements, press
Measure
>
Demod[F2]>AM Analysis[F1]>Min/Max Reset[F3].
AM Pass Fail Testing
The Limit Edit function puts a pass limit on the
AM depth, carrier offset and carrier power.
AM Depth:
1Hz ~ 400kHz
Carr. Offset:
1Hz ~ 400kHz
Carrier Power:
-120dBm ~ 30dBm
1. Press
Measure
> Demod[F2]>AM Analysis[F1]>
Limit Edit[F5] and set the limits.
AM Depth[F1]
If the measured depth is
above this limit, it will be
judged as Fail.
Carr. Offset[F2]
If the measured carrier
offset is above this limit,
it will be judged as Fail.
Carr. Power[F3]
If the measured carrier
power is above this limit,
it will be judged as Fail.
2. Press Pass/Fail and turn Pass/Fail on.