ADVANCED OPERATION
135
9. Press AF Trigger[F5]>Trigger Setup[F1] to set the
triggering conditions:
Edge Slope[F1]
Sets the trigger to rising
or falling edge.
Trigger Mode[F2]
Sets the triggering mode:
Nor.: Normal trigger
Sgl.: Single trigger
Cont.: Continuously
trigger
Trigger Level[F3]
Sets the trigger level as a
percentage of the depth.
(The displayed level will
only remain for a few
moments)
Trigger Delay[F4]
Sets the trigger delay
time:
0 to 1ks
10. Press Return[F6] to return to the AF Trigger
menu and to set the remaining trigger options:
FreeRun[F2]
Disables the trigger, this
is the default setting.
Start Time[F3]
Sets the start time for the
x-axis for the AM
waveform in the top-half
of the screen.
Stop Time[F4]
Sets the stop time for the
x-axis for the AM
waveform in the top-half
of the screen.
Action Now[F6]
Turns FreeRun mode off
and uses the user-
defined trigger settings.