ADVANCED OPERATION 
141 
Disables the trigger, this 
is the default setting.  
Sets the start time for the 
x-axis for the FM 
waveform in the top-half 
of the screen. 
Sets the stop time for the 
x-axis for the FM 
waveform in the top-half 
of the screen. 
Turns FreeRun mode off 
and uses the user-
defined trigger settings. 
The MAX and MIN measurements are held until 
higher or lower values are found. To reset the MAX 
and MIN measurements, press 
>  
Demod[F2]>FM Analysis[F1]>Min/Max Reset[F3]. 
 
FM Pass Fail Testing 
 
The Limit Edit function puts a pass limit on the 
FM deviation, carrier offset and carrier power.  
> Demod[F2]>FM 
Analysis[F2]>Limit Edit[F5] and set the limits. 
If the measured 
deviation is above this 
limit, it will be judged as 
Fail.