ADVANCED OPERATION
221
Spectrum display
Limit Line Table
3. Press Point Num[F1] and choose a point number
to edit with the number pad (must start at #1).
4. Press Frequency[F2] and set the frequency of the
point.
5. Press Limit[F3] and set the amplitude level of
the point.
All the points will be displayed in a limit line
table at the bottom of the display.
6. Repeat steps 3-5 for the remaining points
(maximum of 10points. Points can only be
created in numerical order).
7. To delete the selected point, press Delete
Point[F6].
8. Press Return[F7]>Save Limit Line[F5] to save the
currently selected limit line.
Note
Note that the limit lines are automatically sorted by
frequency (low high).