EasyManua.ls Logo

GW Instek GSP-9300 - Page 331

GW Instek GSP-9300
342 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
APPENDIX
329
Absolute Amplitude Accuracy
Absolute Point
Center=160 MHz ; RBW 10 kHz; VBW 1 kHz;
span 100 kHz; log scale; 1 dB/div; peak
detector; 23 C; Signal at Reference Level
Preamp off
± 0.3 dB
Ref level 0 dBm;
10 dB RF attenuation
Preamp on
± 0.4 dB
Ref level -30 dBm;
0 dB RF attenuation
Frequency Response
Preamp off
Attenuation: 10 dB; Reference: 160 MHz; 20 to
30°C
100 kHz to 2.0
GHz
± 0.5 dB
2GHz to 3 GHz
± 0.7 dB
Preamp on
Attenuation: 0 dB; Reference: 160 MHz; 20 to
30°C
1 MHz to 2 GHz
± 0.6 dB
2 GHz to 3 GHz
± 0.8 dB
Attenuation Switching Uncertainty
Attenuator setting
0 to 50 dB in 1 dB step
Uncertainty
± 0.25 dB
reference: 160 MHz,
10dB attenuation
RBW Filter Switching Uncertainty
1 Hz to 1 MHz
± 0.25 dB
reference : 10 kHz
RBW
Level Measurement Uncertainty
Overall Amplitude
Accuracy
± 1.5 dB
20 to 30°C; frequency > 1 MHz;
Signal input 0 to -50 dBm;
Reference level 0 to -50 dBm;
Input attenuation 10 dB;
RBW 1 kHz; VBW 1 kHz; after cal;
Preamp Off
± 0.5 dB
Typical
±

Table of Contents

Other manuals for GW Instek GSP-9300