9.1 Troubleshooting
163
9
Replaceable Parts
Certain parts require replacement periodically and at the end of their
useful life: (Useful life depends on the operating environment and fre-
quency of use. Operation cannot be guaranteed beyond the following
periods)
The instrument contains a built-in backup lithium battery, which offers a ser-
vice life of about ten years. If the date and time deviate substantially when the
i
nstrument is switched on, it is the time to replace that battery. Contact your
dealer or Hioki representative.
The fuse is housed in the power unit of the instrument. If the power does not
tu
rn on, the fuse may be blown. If this occurs, a replacement or repair cannot
be performed by customers. Please contact your dealer or Hioki representa-
tive.
The instrument contains many electrolytic capacitors. The useful life of electro-
lytic capacitors varies greatly according to the operating environment, so they
may
need to be replaced periodically.
Maintenace and
Service
Chapter 9
9.1 Troubleshooting
Inspection and Repair
Never modify the instrument. Only Hioki service engineers should dis-
assemble or repair the instrument. Failure to observe these precautions
m
ay result in fire, electric shock, or injury.
If damage is suspected, check the "Before returning for repair" (p. 164)"
section before contacting your dealer or Hioki representative.
Transporting
Use the original packing materials when transporting the instrument, if
possible.
Pack the instrument so that it will not sustain damage during shipping,
and include a description of existing damage. We cannot accept
responsibility for damage incurred during shipping.
Part Life
Lithium Battery Approx. 10 years
Backlight
(to half brightness)
Approx. 36,000 hours
Battery pack Approx. 1 year
Electrolytic Capacitors Deterioration occurs after about 10 years when used in se-
vere environments (temperatures around 40°C).