9.4 Disposing of the Instrument
167
9
Before Disposing of the Instrument
The instrument contains a lithium battery for memory backup. Remove
this battery before disposing of the instrument.
9.4 Disposing of the Instrument
• To avoid electrocution, turn off the power switch and disconnect
the power cord and cables before removing the lithium battery.
• Handle and dispose of batteries in accordance with local regula-
tions.
Removing the Lithium Battery
Required tools:
• One Philips screwdriver (to remove screws)
• One tweezers (to remove the lithium battery)
1
Turn the instrument over to re-
move the retaining screw from
the battery pack compartment
cover, and remove the cover.
2
Remove the six screws affixing
the two halves of the case.
To remove the battery, insert the twee-
zers between the lithium battery holder
and the battery to grasp the battery.
3
CALIFORNIA, USA ONLY
This product contains a CR Coin Lithium Battery which contains
Perchlorate Material - special handling may apply.
See www.dtsc.ca.gov/hazardouswaste/perchlorate