4.2 Selecting the Measurement Method (SYSTEM Screen) [MEASURE]
60
Select the model name of the clamp sensors to be used.
This setting is ignored if no clamp sensors are used.
For the detail about the clamp sensor, refer to the instruction manual pro-
vided with the device.
Selecting the Clamp Sensor Model and Current Range
Move to the setting item.
Select the screen.
1
2
Select from the pull-down
menu.
3
Accept
Cancel
Clamp sensor Current range
9660
10A/ 100A 9660 Clamp-on Sensor
9661
50A/ 500A 9661 Clamp-on Sensor
9667_500
50A/ 500A 9667 Flexible Clamp-on Sensor (500A range)
9667_5k
500A/ 5000A 9667 Flexible Clamp-on Sensor (5000A range)
9669
100A/1000A 9669 Clamp-on Sensor
9694
5A/ 50A 9694 Clamp-on Sensor
9695-02
5A/ 50A 9695-02 Clamp-on Sensor
9695-03
10A/ 100A 9695-03 Clamp-on Sensor
9657-10
0.5A/ 5A 9657-10 Clamp-on Leak Sensor
9675
0.5A/ 5A 9675 Clamp-on Leak Sensor
The current ranges available for selection depend on the clamp sensor model.
The range setting applies to all channels (1 to 3). Different ranges cannot be set for indi-
vidual channels.
To use non-Hioki clamp sensors, bear in mind the following requirements:
Only voltage-output type clamp sensors can be used.
Output should not exceed 1.7 Vrms (2.4 Vpeak).
The ratio of output voltage to input current should be the same as Hioki clamp sensors.
Refer to "Appendix 5 Power Range Structure" (p. A9).
When the clamp sensor model setting is changed, the highest current range is ini-
tially selected.