11.3 Error Indication
180
Contact your dealer (agent) or local sales office if a repair should become necessary.
Different sensors! Cannot
change the wiring in the set-
ting file.
Attempted to load an incompatible setting
configuration file.
Settings cannot be loaded unless instrument
options and saved items are the same as those
installed and selected when saved.
See "7.9 Reloading Setting Configurations"
(p. 116)
D/A option is different.
Attempted to load an incompatible setting
configuration file.
Motor option is different.
Attempted to load an incompatible setting
configuration file.
Inconsistent items to save
Attempted to load an incompatible setting
configuration file.
CF card error!
This card is not supprted.
Incompatible CF card found.
Use a Hioki CF card option.
See "Chapter 7 Data Saving and File Opera-
tions" (p. 103)
USB memory stick error!
This memory stick is not sup-
prted.
Incompatible CF card found.
Use a Hioki CF card option.
See "Chapter 7 Data Saving and File Opera-
tions" (p. 103)
Failed to write. Writing to storage media failed.
Try again.
Failed to read. Reading from storage media failed.
Failed to save while calculat-
ing the waveform data
Attempted to save a waveform while it is
being created.
Try again after the waveform is created (when
the timer mark disappears).
Failed to create a file. File creation failed for unknown reason.
Try again.
Failed to create a folder. Folder creation failed for unknown reason.
Printer error!
Attempted printer auto-setup without a
printer connected, or with printer turned off.
Check that the printer is connected and turned on.
See "8.1.1 Printer Preparation and Connec-
tion" (p. 124)
Unknown error! An unknown error has occurred.
Clear this error by pressing any key except
or once. If the error recurs, contact your
dealer or Hioki representative.
Error display Cause Remedy
If the measurement lines are energized when the instrument is turned on, the instrument may
be damaged or an error message may appear, so before energizing the lines, turn the instru-
ment on and confirm that no error message is displayed.