4.8 Viewing Motor Measurement Values (With Hioki 9791 or 9793 installed)
88
When [CHB input] is set to [Pulses], make these five settings to suit the rotation signal: [CHB unit],
[Max frequency], [No. of pulses], [Motor poles], and [CHZ].
When [Pulses] is selected
CHB unit Always select [r/min] when measuring motor power (Pm).
Hz, r/min
The measurement range when CH B units are set to Hz is 0.5 Hz to 5 kHz.
Measurement values are calculated as
When the pulse signal input frequency is higher than the measurement range, set the
appropriate pulse count.
No. of Poles×Pulse Input Frequency
2 ×Pulse Count Setting
Max frequency Determine the full-scale measurement value for CH B.
(When CH B is set for D/A Output, this setting is the full-scale value.)
100 Hz, 500 Hz, 1 kHz, 5 kHz
Pulse Count Set the number of pulses per mechanical rotation, from 1 to 60,000.
Available setting values are multiples of half the number of motor poles.
(For a incremental type rotary encoder that provides 1000 pulses per rotation, set to 1000.)
+1/2 the number of motor poles,
-1/2 the number of motor poles
Increment or decrement by 1/2 the number
of motor poles.
+1/2 × 10 times the number of motor poles,
-1/2
× 10 times the number of motor poles
Increment or decrement by 1/2 × 10 times
the number of motor poles.
+1/2 × 100 times the number of motor poles,
-1/2
× 100 times the number of motor poles
Increment or decrement by 1/2 × 100 times
the number of motor poles.
No. of Motor
Poles
Set the number of motor poles as an even number between 2 and 98.
(The slip calculation and the rotation signal input as the frequency corresponding to
the mechanical rotation angle are converted to the frequency corresponding to the
electrical angle.)
+2, -2 Increment or decrement by 2.
+10, -10 Increment or decrement by 10.
The motor pole setting is enabled by pressing (Set). Be sure to press (Set)
after changing the setting.
When using pulse synchronization with an electrical angle (motor input voltage, or the same
frequency as the current frequency), set the number of motor poles at CH B input to 2.
CHZ Make CH Z settings.
Enable (set to ON) to use the original rotation angle (usually called the Z phase) signal
to reset pulse division to zero.
ON Use CH Z.
OFF Ignore CH Z (no connection to CH Z jack)