EasyManua.ls Logo

Hitachi TM3030 Plus - Page 187

Hitachi TM3030 Plus
384 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 - 68
1a. Low mag.
This observation condition tends to produce smooth images with high contrast,
suitable for observation at a magnification of 15x to 2,000x due to low resolution. This
magnification is only reference value. Set the advanced value depending on samples
and purpose.
When [Low mag.] is selected, the [l] mark is displayed in the information area located
in the lower right part of the image observation area. (see 4.3.2)
Figure 4.3.12-2 Observation Condition Settings (5 kV - Low Mag.)
1b. Middle mag.
An electron beam is focused more than [Low mag.], achieving relatively high
resolution. However, in some cases, contrast or image degradation can occur
depending on the observed sample or magnification. This condition is suitable for
observation at a magnification of 2,000x to 6,000x. This magnification is only
reference value. Set the advanced value depending on samples and purpose.
When [Middle mag.] is selected, the [m] mark is displayed in the information area
located in the lower right part of the image observation area. (see 4.3.2)
Figure 4.3.12-3 Observation Condition Settings (5 kV - Middle Mag.)
1c. High mag.
An electron beam is more focused than [Middle mag.], achieving high resolution.
Although charge-up phenomenon and damage to samples can be minimized,
contrast or image degradation can occur depending on the observed sample or
magnification. This condition is suitable for observation at a magnification of 6,000x to
30,000x. This magnification is only reference value. Set the advanced value
depending on samples and purpose.
When [High mag.] is selected, the [h] mark is displayed in the information area
located in the lower right part of the image observation area. (see 4.3.2)

Table of Contents

Related product manuals