Page 170 - CEAST 9340 - Instructions for Use and Maintenance 7510.000MN1r ed. 1 rev. 1
List of Error Codes (Input Parameters)
1 of 2
Code Error Cause Remedy
100 Wrong parameter Parameter input error Type-in the correct parameter
101 Insufficient impact speed Parameter input error. Velocity
value too low
Increase the velocity value
102 Excessive drop height Parameter input error. Drop
height value too high
Decrease the drop height
value
103 Insufficient drop height Parameter input error. Drop
height value too low with the
antirebound enabled
Increase the drop height value
104 Excessive impact speed Parameter input error. Velocity
too high
Decrease the velocity value
105 Temperature range set
between 0 and 40 °C
Parameter input error.
Temperature value out of range
Type-in the temperature value
within ranges: - 70 to 0°C and
40 to 150 °C
106 Set temperature is lower than
limit
Parameter input error Increase the temperature
value. Lower limit is - 70 °C
107 Set temperature is higher than
limit
Parameter input error Decrease the temperature
parameter. Upper limit is 150
°C
108 Temperature input error Parameter input error Type-in the correct value
109 Date input error Parameter input error. Date
and hour setting error
Type-in the correct date and
hour
110 Incorrect input of identifying bit
of the base
Parameter input error: wrong
instrument configuration
Type-in the bit that identify the
correct base
111 ParError: lot number not
correct
Parameter input error: wrong
lot number
Type-in the correct number
112 ParError: new password not
correct
Parameter input error: wrong
password
Type-in a correct password
(use letters and numbers only,
8 characters max.)
113 ParError: password level not
correct
Parameter input error: wrong
password level
Type-in the correct level
114 ParError: language code not
correct
Parameter input error: wrong
language or not existing one
Select another language