Page 172 - CEAST 9340 - Instructions for Use and Maintenance 7510.000MN1r ed. 1 rev. 1
List of Error Codes (Emergency Warnings)
1 of 2
Code Error Cause Remedy
200 Upper door open Door of the turret leaved open Close the door
202 Current overload on crosshead
motor
Safety intervention for current
overload on striker lifting
crosshead motor. It can also
appear if the tup is stuck after
penetration of a specimen, try
to move up and down the lifter
or to remove by hand or other
tool the specimen from the tup
if the tup is stuck after
penetration of a specimen, try
to move up and down the lifter
or to remove by hand or other
tool the specimen from the tup.
otherwise contact Instron
Service
203 Emergency button or power off Safety intervention. Emergency
button leaved pressed or
power off
Restore the correct operating
condition
205 Chamber door open Safety intervention. Test
chamber door open
Close the door
207 Thermoregulator error (T out of
range)
Temperature regulator of the
environmental chamber.
Temperature value higher than
the upper limit
Contact Authorized Service
209 Thermoregulator emergency Safety thermostat intervention.
Temperature regulator fault
Contact Authorized Service
210 Error handler due an
emergency status
Actuator emergency Check emergency condition:
door open, power off, low air
pressure, emergency button
pressed, .......
211 Error handler due an
emergency status (Striker)
Emergency condition occurred
during the striker movement
Check emergency condition:
door open, power off, low air
pressure, emergency button
pressed, .......