CEAST 9340 - Instructions for Use and Maintenance 7510.000MN1r ed. 1 rev. 1 - Page 173
List of Error Codes (Emergency Warnings)
2 of 2
Code Error Cause Remedy
213 Error handler due an
emergency status (Shock
absorbers)
Emergency condition occurred
during the antirebound
movement
Check emergency condition:
door open, power off, air
pressure, .......
214 Error handler due an
emergency status (Specimen)
Emergency condition occurred
during the specimen
movement
Check emergency condition:
door open, power off, air
pressure, .......
215 Error handler due an
emergency status (Shutter)
Emergency condition occurred
during the striker releasing
command in the test cycle.
Shutter remains closed
Check emergency condition:
door open, power off, air
pressure, .......
217 Error handler due an
emergency status (Crosshead)
Emergency condition occurred
during the striker lifting
crosshead movement
To reset the inverter leave
open the turret door for 5
seconds.
Check emergency condition:
door open, power off, air
pressure, .......
219 Error handler due an
emergency status (Restart T.)
List of Error Codes (Test Starting Condition) 1 of 1
Code Error Cause Remedy
301 Temperature not ok Environmental chamber
temperature out of tolerance at
the test start
Wait until the temperature
value is reached before start
the test
310 Photocell not ready Impact and rebound velocity
optical detector fault
Contact Authorized Service
311 Photocell not reset Contact Authorized Service