CEAST 9340 - Instructions for Use and Maintenance 7510.000MN1r ed. 1 rev. 1 - Page 53
When the door is closed again, the systems inside the environmental chamber
will be switched-on after a few seconds.
The clamping plate goes down and clamp the films just before that the fan
is switched-on. This condition avoids the expulsion of film from the specimen
support due to the air flow of the fan.
Note: The “I” option is available only with environmental chamber configured.
4) Input H/V/E”: select the input type (H, V or E):
H: drop height;
V: impact velocity;
E: energy at the impact.
5) Conditioning”: it enables (Y) or disables (N) the temperature control of environmen-
tal chamber (if installed and configured in the program). Function not used for stan-
dard chamber.
Conditioning N: the test starts when the operator runs it.
Conditioning Y: the test can be run only if the temperature is at the set point (± 1 °C).
6) Antirebound”: it enables (Y) or disables (N) the striker antirebound system (if in-
stalled and configured in the program).
7) Other param.”: it goes to the next menu of parameters (see Vid. 1.3 b).