Page 80 - CEAST 9340 - Instructions for Use and Maintenance 7510.000MN1r ed. 1 rev. 1
ASSEMBLY
X
a) Open the front door of the turret accessing to the lifting system, using the
special key provided.
b) Place the instrumented striker against the striker holder (Fig. 2, detail A) cen-
tering the existing pins and making sure that the signal cable is facing to the
front of the instrument (check not required when install a striker not instru-
mented).
c) Fasten the striker with screws.
d) Insert the signal cable in the three harmonic wire eyelets (Fig. 2).
e) Extend the signal cable following the route shown in Fig. 2 and fasten the cable
holder with the clamp (Fig. 2, detail B) pushing it upward.
f) Continue with the signal cable downward and insert it in the slot outside (Fig. 2,
detail C) on the left side of the turret.
g) Fasten the signal cable to the structure using the clamp (Fig. 2, detail D).
h) Screw the connector of the signal cable on the rear panel of the DAS (Data
Acquisition System, see para. 7.1 for connection).
Notes: - Make sure that the signal cable of the striker is long enough to guarantee the
striker stroke up to the maximum height.
- Check that the striker signal cable is far enough from the cylinder (on the left side
of turret) of antirebound system avoiding to cut the cable during the movement
of the antirebound crosshead or during the striker repositioning.
X
X
X
X
X
X
X