Chapter 9: Qualification Study
ValProbe RT User Manual 94
defined in the Setup to occur automatically based on date / time or Cycle Time, then the
button is grayed out and occurs automatically when the condition occurs.
At the end of the Qualification study, select the Save Study button to save the study to the
Asset under Qualifications.
All actions are documented in the Messages pane and Audit Trail. The Quit Qual button in the
upper right of the display, allows a User to quit a study in progress. The User is prompted to
save the study. The Qualification file contains all data up to the time the study was aborted.
9.4. Viewing Qualification Real-Time Displays
During a Qualification Study the ValProbe RT software offers numerous real-time and
historical displays to easily evaluate and analyze the performance of the Qualification study.
The displays include List, Graph, Layout, and Historical views. All views contain similar
structure and elements:
• A Header including the Asset name, Setup Name, and IP address of the connected
Base Station.
• A display windows which switches between List, Graph, and Layout view depending
the selection on the top right of screen.
• An Events table listing events for all Groups
• A Calculation table displaying real-time Statistical Calculations for the selected Group.
Values listed include
• Min
• Min-Max
• Average
• Max Average
• Average Min
• CycleDuration: How long study has been running
• ExposureDuration: How long exposure has been running
• Battery %
• RF Link
• Sample %: the percentage of data in the Logger that has been transmitted into the
Base Station
• ALethality
• Control Buttons to control the Start / Stop of Qualification and Exposure.
• A Setup button to review the parameters configured in the Setup.
• A Hardware button to provide details of connected devices.
• A Quit Qual button to cancel the Qualification Study and delete all data samples.
• An Add User Event button to mark additional events during a running study.