EasyManua.ls Logo

Kaye ValProbe RT - Data Collection

Kaye ValProbe RT
181 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2: Kaye ValProbe RT Hardware
ValProbe RT User Manual 6
Chapter 2. Kaye ValProbe RT Hardware
2.1. Kaye ValProbe RT Loggers
The ValProbe RT loggers measure three critical measurands and in eight different hardware
configurations:
Temperature - Measures temperature from 85°C to 140°C. The bendable Temperature
Logger may be used in conjunction with an Insulating Canister to measure temperature
up to 400°C.
Pressure and Temperature - Measures pressure from 0 to 5bar absolute and
temperature from 0°C to 140°C.
Figure 5: Kaye ValProbe RT Loggers
The Kaye ValProbe RT Loggers are made from pharmaceutical grade polyether ether ketone
(PEEK), 316 stainless steel, and designed for in-situ use. For example, Loggers immersed in
cleaning fluid or detergents, or used in harsh environments. Loggers store study setups and
up to 100,000 pressure samples and 100,000 temperature samples with variable stop / start
conditions in non-volatile memory.
WARNING! Substitution of components may damage the Logger.
Do not heat the Temperature Logger above 140°C.
Do not heat the Pressure/Temperature Logger above 140°C.
Do not submerge the Logger in oil, cleansing liquid or detergents for more
than two hours at temperatures below 20°C. After two hours the O-ring
seal may degrade and should be replaced.
Do not heat the Humidity/Temperature Logger with RH sensor above 70°C.
ATTENTION ! La substitution de composants peut endommager l'enregistreur.
Ne chauffez pas l'enregistreur de température à plus de 140°C.
Ne pas chauffer l'enregistreur de pression/température à plus de 140°C.

Table of Contents