Chapter 11: Sensor Verification
ValProbe RT User Manual 110
11.4. Loading a Humidity Verification Setup
Before beginning a Humidity Verification process, load a setup into the ValProbe RT Loggers.
A Humidity Chamber must be used to execute a Verification study. Please connect it to the
Base Station connectors on the rear of the Base Station. The connector is marked with the
appropriate icon. Utilize the supplied cables for the connection. Power up the Humidity
Chamber.
To load a setup:
From the Main Menu navigate to the Equipment-> Setup-> Humidity / Temperature
Verification and select a setup from list of available setups.
After highlighting the required Verification Setup, click Select Base station.
Figure 145: Selecting a Humidity Setup
On the Select Base Station screen, select Discover Base Station, available devices are
displayed. If there is an issue during discovery, ensure that the Base Station and Validation
Console are properly connected
If no Base Stations are displayed check the connections and settings:
If using direct connection using fixed IP, ensure the supplied ethernet cable is
connected directly from the Base Station to the Console docking
If the System is on the network (DHCP), ensure the Console and Base Station are
connected via ethernet cables to the network via ethernet connections.