Appendix A. Specifications
ValProbe RT User Manual 151
A.3.11 Insulating Canister X2545
Size 149mm (5.87”) x 45mm (1.75”) (Length x diameter), weight 330grams (0.73 lb).
A.3.12 Battery
Field replaceable 3.6VDC ½ AA Lithium Thionyl Chloride primary battery.
A.3.13 Shelf Life
> ten years.
A.3.14 Battery Life
Battery life is affected by several factors, including sample rate, transmission rate,
temperature, Logger type and hours of use. Individual battery capacity can vary from one
battery to another by up to 10%.
A.4 Kaye ValProbe RT Temperature Logger
A.4.1 Measurement Accuracy
±0.1°C from -85 to 140°C.
±0.25°C from 140 to 400°C for bendable models.
A.4.2 Body Environmental Ratings:
Temperature: -85 to 140°C.
Temperature: -85 to 360°C with canister model.
Humidity: 0 to 100%, condensing.
Pressure: 0 to 5 bar absolute.
A.4.3 Number of Input Channels
1, 2 and 5 channels (5ch Logger sampling rate is min two secs).
A.4.4 Sensor Types and Measurement Range
Rigid, flexible, bendable and surface mount: -85 to 140°C.
Bendable Loggers: till 400°C (on request – please contact factory).
A.4.5 Sensor Lengths
Rigid 1.5, 3, 6 and 9” inches. Bendable 12, 24 and 36” inches. Flexible 40” inches.
Please contact factory for special lengths.