Instrument Messages
Event Messages
–221 Settings conflict;
Destination trace for Trace
Math cannot be a trace
operand
The resulting trace data (from doing a trace math function)
cannot be put into the any of the traces that are being used
by the math operation.
–221 Settings conflict; Downconv
only available when DUT is
Amplifier
SCPI only message. The System Downconverter can only be
set to ‘On’ when the DUT type is amplifier. Change DUT type
to Amplifier if the System Downconverter is required.
–221 Settings conflict; EDGE EVM
only supports EDGE TCH burst
type.
–221 Settings conflict;
Electronic attenuator is
disabled
You are using the mechanical attenuator, and have not
enabled the electronic attenuator. You cannot set the value
of the electronic attenuator because it automatically
sets/changes when enabled.
–221 Settings conflict;
Electronic attenuator is not
available above 3.6 GHz
The maximum frequency of the electronic attenuator is 3.6
GHz. This is because of switching capacitance.
–221 Settings conflict;
Electronic attenuator
unavailable in current state
–221 Settings conflict;
Electronic attenuator
unavailable with Preamp on
The internal preamp is on. Electronic attenuator cannot be
used while you are using the internal preamp.
–221 Settings Conflict; FAST
method can only be used
while Radio Std is W-CDMA
–221 Settings conflict; Feature
not available in this View
Some functionality is available in one View, but not in
another. (See the Views under the View/Display key.) This
error occurs if you send a SCPI command or push a gray-out
key that is not available in the current selected View.
–221 Settings conflict; Feature
not supported for selected
source
You have asked for a feature that the selected source is not
capable of.
–221 Settings conflict; Feature
not supported for this
measurement.
Some functionality is available in one measurement, but not
in another. (See the measurements under the Meas key.)
This error occurs if you send a SCPI command or push a
gray-out key that is not available in the current selected
measurement.
–221 Settings conflict; Feature
not supported for this model
number
This functionality is not a part of the instrument you are
using but may be found in other models in the X-Series.
Err# Message Verbose/Correction Information