EasyManuals Logo

Keysight Technologies N9030B User Manual

Keysight Technologies N9030B
617 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #537 background imageLoading...
Page #537 background image
N9030B PXA Signal Analyzer Service Guide 537
Assembly Replacement Procedures
Input Connector Assembly
Input Connector Assembly
Removal
1. Remove the instrument outer case. Refer to the Instrument Outer Case
removal procedure.
2. Remove the Front Frame Assembly. Refer to the Front Frame Assembly
removal procedure.
3. Refer to Figure 16-97. Disconnect the semi-rigid cable W2 from the Input
Connector Assembly.
Figure 16-97 Input Connector Assembly Removal
4. Remove the two screws (1) (0515-0372) that attach the Input Connector
Assembly to the chassis. The connector assembly can now be lifted from
the chassis.
Replacement
1. Refer to Figure 16-97. Place the Input Connector Assembly into position in
the chassis. Replace the two screws (1) (0515-0372) to attach the Input
Connector Assembly to the chassis. Torque to 9 inch-pounds.

Table of Contents

Other manuals for Keysight Technologies N9030B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N9030B and is the answer not in the manual?

Keysight Technologies N9030B Specifications

General IconGeneral
BrandKeysight Technologies
ModelN9030B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals