216 N9030B PXA Signal Analyzer Service Guide
Front End Control Troubleshooting
A15 Front End Control Assembly Troubleshooting
Verifying SW3 (Option LNP only)
Since there is a ribbon cable from the A15J802 to SW3, it is not feasible to use
a voltmeter or oscilloscope to measure the control voltages. For testing SW3,
the “click” test can be used.
Press Mode Preset on the analyzer. Press FREQ, 5 GHz, SPAN, 1 MHz. Verify
the optional preamplifier is off by pressing AMPTD, More 1 of 2. The Internal
Preamp softkey must be Off in order to test SW3. Press AMPTD, More 1 of 2,
uW Path Ctrl, Low Noise Path Enable on the analyzer. You should hear a
distinct “click” noise from SW3 when the low noise path is selected. Now press
uW Path Ctrl and select Standard Path. SW3 will “click” again and return to
the standard RF path.
Another way of testing this switch is using an RF signal. This is described in the
RF Section Troubleshooting chapter, under the high band troubleshooting
section.
Verifying SW1 and SW2 (Option MPB only)
SW1 and SW2 can be verified three different ways:
—“Click” Test
— Oscilloscope
— R.F. Signal Tracing
“Click” Test
This test changes between two different path modes. When this is done a
distinct “click” noise from SW1 and SW2 can be heard. Since both switches
switch at the same time, it is difficult to decipher if one is switching and the
other is not.
Press Mode, Preset on the analyzer. Press FREQ, 5 GHz, SPAN, 1 MHz,
AMPTD, More 1 of 2, uW Path Ctrl. Standard Path is the default. Select uW
Preselector Bypass and you should hear a distinct “click” noise if the control
logic from the A15 Front End Control board is getting to SW1 and SW2.