EasyManuals Logo

Keysight Technologies N9030B User Manual

Keysight Technologies N9030B
617 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #96 background imageLoading...
Page #96 background image
96 N9030B PXA Signal Analyzer Service Guide
Instrument Messages
Event Messages
–300 to –399, Device-Specific Errors
–440 Query UNTERMINATED after
indefinite response
A query was received in the same program message after a query
requesting an indefinite response was executed.
Err# Message Verbose/Correction Information
–300 Device-specific error An instrument error occurred and the exact problem cannot be specifically
identified. Report this error to the nearest Keysight Technologies sales or
service office.
–310 System error; An internal system-type error has occurred. The exact problem cannot be
specifically identified. Report this error to the nearest Keysight
Technologies sales or service office.
–310 System error; A
license will soon
expire;<feature code>
will expire in <time>
The indicated feature/software will expire in the specified time. Contact
Keysight Technologies to purchase continued use of this functionality.
–310 System Error; enable
GPIB controller mode
Press System, I/O Config, GPIB and set GPIB Controller to Enabled so that
the analyzer can control the source over GPIB
–310 System error; Error
transmitting a LAN
event to the network.
Communication with the network driver failed.
–310 System error; Failed
to initialize the PTP
clock to current
time.
Failure communicating with the DMC libraries’ PTP controller.
–310 System error; Failed
to instantiate the
PTP ordinary clock.
Failure in the starting up the DMC libraries’ PTP controller.
–310 System error; feature
<feature code> not
licensed
The specified feature, for example “N9073A-TR2” is not licensed. The
license may have expired. You cannot use it until you get a license.
–310 System error; Feature
expired; <feature
code>
The specified feature has expired. The license is no longer valid.
–310 System error; License
installation
failed;<feature code>
The license installation of the specified feature, for example
“N9073A-TR2”, has failed. You should refer to the event log in the control
panel for more details.
–310 System error; License
removal failed;
<feature code>
The license removal of the specified feature, for example “N9073A-TR2”
has failed. You should refer to the event log in the control panel for more
details.
Err# Message Verbose/Correction Information

Table of Contents

Other manuals for Keysight Technologies N9030B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N9030B and is the answer not in the manual?

Keysight Technologies N9030B Specifications

General IconGeneral
BrandKeysight Technologies
ModelN9030B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals