230 N9030B PXA Signal Analyzer Service Guide
Analog/Digital IF Troubleshooting
A2 Analog IF Troubleshooting
A2 Analog IF Troubleshooting
There are three steps to verify the A2 Analog IF Assembly.
Measure the input power and frequency accuracy of the 322.5 MHz signal
(from the A13 RF Front End assembly)
Measure the output power and frequency accuracy of the 22.5 MHz signal
(to the A3 Digital IF Assembly)
Measure the 300 MHz input LO signal (from the A16 Reference Assembly)
Verifying the 322.5 MHz Input Power
1. Turn the instrument off.
2. Remove instrument cover. Refer to Chapter 16, “Assembly Replacement
Procedures” in this service guide.
3. Turn on the PXA Signal Analyzer and wait for the instrument to complete
the boot up process.
4. Press System, Alignments, Auto Align, Off.
5. Press Input/Output, RF calibrator, 50 MHz.
6. Verify the 50 MHz signal is at 25 dBm on the analyzer display by pressing
FREQ, 50 MHz, SPAN, 1 MHz, Peak Search on the PXA. The marker
readout should be 50 MHz at 25 dBm ± 3 dB. If this reference signal is
measuring incorrectly, see Chapter 4, “RF Section Troubleshooting
(RF/Microwave Analyzers)” in this service guide.
7. To continue verifying press SPAN, Zero Span. Verify the input attenuator
on the PXA is set to 10 dB. Look near the top of the display near the center
and verify that Atten: 10 dB is visible. If needed change the input
attenuator by pressing AMPTD, Attenuation, 10 dB on the analyzer.

Table of Contents

Other manuals for Keysight Technologies N9030B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N9030B and is the answer not in the manual?

Keysight Technologies N9030B Specifications

General IconGeneral
BrandKeysight Technologies
ModelN9030B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals