326 N9030B PXA Signal Analyzer Service Guide
Hardware Options
A25 Wideband Analog IF Troubleshooting - Option B1X and B85
Inputs to the A25 Assembly
A25J101 300 MHz IF signal (from A15J901 Front End Controller)
A25J301 2400 MHz reference from A16 Reference Assembly via W34 cable
A25J805 Step Calibrator from A3J17 40 MHz Digital IF via W45 cable
Outputs from the A25 Assembly
A25J302 2400 MHz Reference Out
A25J102 250 MHz IF Out to A3J15 of the 40 MHz Digital IF
A25J806 50 MHz Comb Out to A16 Reference assembly A16J726 via W46
cable
A25J2 ADC Data to 160 MHz Digital IF A26J504 via W44 cable
A25 Wideband Analog IF Assembly Theory of Operation - Option B1X
and B85
The A25 Wideband Analog IF assembly primary role is to properly filter and
apply attenuation or gain to optimize the level going to the ADC and provide
sufficient pre-filtering to reject images that would otherwise interfere with the
signal of interest. Precision alignments are used to calibrate the filters,
attenuators and gain of this assembly. It is not uncommon that the reported
failure will come from failures reported when an alignment fails for this
assembly. The alignment history file is a good indicator of a failure for this
assembly. This file is a text file that can be edited with Notepad. The file is
located in the following location:
E:\AlignDataStore\AlignmentHistory.txt
This file will list all alignments that have passed or failed. Alignments reporting
a failure of the WBAIF assembly can in many cases be attributed to failure of
this assembly. If however these errors follow errors in the RF section of the
instrument, it is possible that additional WBAIF error will appear due to
missing RF Calibration signals and not related to the A25 Wideband IF
assembly.
Manually troubleshoot the assembly is basically verifying the 300 MHz IF signal
is present and at the proper level. In addition you will want to verify input and
output levels of the various calibration and reference signals.
Additionally, the A25 Analog IF has a built in RF Burst Trigger.

Table of Contents

Other manuals for Keysight Technologies N9030B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N9030B and is the answer not in the manual?

Keysight Technologies N9030B Specifications

General IconGeneral
BrandKeysight Technologies
ModelN9030B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals