N9030B PXA Signal Analyzer Service Guide 143
Instrument Messages
Condition Messages
Condition Errors 601 to 699, Error Summaries
This series of errors corresponds to the bits in the STATus:QUEStionable
register, read with a STATus:QUEStionable? event query or a
STATus:QUEStionable:CONDition? query.
The second column in the table shows the corresponding bit in the status
register. These bits do not have any corresponding error messages, although
error numbers have been reserved for them as seen in the Err# column; they
are status bits only.
The bits in the STATus:QUEStionable register are “summary bits” for registers at
a lower level. Note that these summary bits summarize the state and history of
the event registers at the lower level. This is true even for bits in the
STATus:QUEStionable condition register. This means that:
The summary bits read by the STATus:QUEStionable: CONDition? query are
true if any event bits are set in any of the :QUEStionable sub-registers :POWer,
:TEMPerature, FREQuency, CALibration or :INTegrity.
The summary bits read by the STATus:QUEStionable? event query are true if
any event bit has undergone a false-to-true transition with the PTRansition
filter set, or a true-to-false transition with the NTRansition filter set, in any of
the :QUEStionable sub-registers :POWer, :TEMPerature, FREQuency,
CALibration or :INTegrity.
Thus, the summary bits cannot be used to determine the current state of a
lower level condition bit; only the state and history of the lower level event bits.
Err# Bit in
status
register
Message Error or
Warning
More Information
601 0 Voltage unused
603 1 Current unused
605 2 Time unused
607 3 Power status bit
only
This bit is the summary bit for the
STATus:QUEStionable:POWer sub-register.
609 4 Temperature status bit
only
This bit is the summary bit for the
STATus:QUEStionable:TEMPerature
sub-register.
611 5 Frequency status bit
only
This bit is the summary bit for the
STATus:QUEStionable:FREQuency
sub-register.
613 6 Phase unused
615 7 Modulation unused

Table of Contents

Other manuals for Keysight Technologies N9030B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N9030B and is the answer not in the manual?

Keysight Technologies N9030B Specifications

General IconGeneral
BrandKeysight Technologies
ModelN9030B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals