N9030B PXA Signal Analyzer Service Guide 99
Instrument Messages
Event Messages
–221 Settings Conflict Errors
This is one of the errors in the standard SCPI error range of –200 to –299. See
the table “–200 to –299, Execution Errors” on page 110.
The <subtext> part of a Settings Conflict error should be worded so that the
text is: “function1” is not whatever/with/while/when “function2”. This makes
them easier to find - alphabetically, to avoid duplicates.
The entire message displays in the error history as “–221, Settings conflict;
<subtext>”
For example, –221.0076 displays as:
–221, Settings conflict; Invalid trace number
Err# Message Verbose/Correction Information
–221 De-emphasis only available
in FM
The de-emphasis function is only available if FM demod is
selected.
–221 Function not available in
Zero Span
The function you are trying to access is not available in zero
span.
–221 Setting conflict; <trigger
source> trigger is not
available while input is
<input port>
The trigger source (Video, RF Burst, I/Q Mag, etc.) is not
available with the current input port (RF, IQ, etc.)
–221 Setting conflict;
Differential setting
determined by probe type
A probe is connected that has a built in Differential setting.
The setting cannot be changed manually.
–221 Setting conflict; Input Z
unavailable when probe
sensed
A probe is connected and the Input Z is set based on the
probe type. It cannot be changed manually.
–221 Settings conflict; A legal command was received but it could not be executed
due to the current device state.
–221 Settings conflict; *.CSV
file format is not available
in this measurement.
You cannot load or save base instrument traces, as this is not
supported by the Log Plot measurement.
–221 Settings conflict; <Q Param>
cannot be changed when Q
same as I
When the “Q Same as I” parameter is set to Yes, the I
parameter value is copied to <Q Param> and the <Q Param>
value cannot be changed. Set Q Same as I to No to enable
explicit control of the <Q Param> value.
–221 Settings conflict; A Valid
User Cal is required.
Optimize aborted
Optimize Preselector can only be performed if a valid user
cal exists and is applied to current results. Perform a user
cal first or apply existing cal.
–221 Settings conflict; Auto Tune
not available in Tracking
Source mode
The Auto Tune feature cannot be used when you are using a
Tracking Source.