98 N9030B PXA Signal Analyzer Service Guide
Instrument Messages
Event Messages
–312 PUD memory lost Protected user data saved by the *PUD command has been lost.
–313 Calibration memory
lost
The nonvolatile calibration data used by the *CAL? command has been lost.
–314 Save/recall memory
lost
The nonvolatile data saved by the *SAV? command has been lost.
–315 Configuration memory
lost
The nonvolatile configuration data saved by the instrument has been lost.
–320 Storage fault; A problem was found while using data storage. The error is not an
indication of physical damage or failure of any mass storage element.
–321 Out of memory An internal operation needed more memory than was available. Report this
error to the nearest Keysight Technologies sales or service office.
–330 Self-test failed A self-test failure occurred. Report this error to the nearest Keysight
Technologies sales or service office.
–340 Calibration failed The instrument requires an Align All Now. Restore the alignment by
pressing System, Alignments, Align All Now.
–340 Calibration failed;
<failure msg> <port>
The calibration for one of the I-Q ports did not succeed. The information in
the “failure msg” field can be used to troubleshoot this problem. Contact
Keysight technical support if necessary.
–350 Queue overflow An error occurred that did not get put in the error queue because the
queue was full.
–360 Communication error There was a problem with instrument remote communications. The exact
problem cannot be specifically identified.
–360 Communication error;
SNS data read
failure. Disconnect
then reconnect SNS
The Keysight Smart Noise Source connected to the PXA has failed to be
read by the application. Please disconnect and reconnect the SNS. If this
continues to fail, then the SNS may have had its EEPROM corrupted or
another hardware fault exists. Check SNS on another instrument, NFA and
ESA are also SNS compatible instruments. Check the device is not an
Keysight power sensor which uses the same cable interface.
–361 Parity error in
program message
A parity bit was not correct when the data was received. For example, on a
parallel port.
–362 Framing error in
program message
A stop bit was not detected when data was received. For example, on a
remote bus port.
–363 Input buffer overrun A software or hardware input buffer on a port overflowed with data
because of improper or nonexistent pacing.
–365 Time out error There was a time-out problem in the instrument. The exact problem cannot
be specifically identified.
Err# Message Verbose/Correction Information

Table of Contents

Other manuals for Keysight Technologies N9030B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N9030B and is the answer not in the manual?

Keysight Technologies N9030B Specifications

General IconGeneral
BrandKeysight Technologies
ModelN9030B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals