N9030B PXA Signal Analyzer Service Guide 97
Instrument Messages
Event Messages
–310 System error; No
license; <feature
code> will terminate
in <time>
The specified feature will stop working in the specified time due to the
license expiration You will be prompted to save results and exit.
–310 System Error; No
supported source
found
Signal source at given IP address is not responding / IP does not belong to
a source. Check IP address and network connection.
–310 System Error; source
connection lost,
check interface
connection
Signal source at given IP address is not responding / IP does not belong to
a source. Check IP address and network connection.
–310 System error; The
configured PTP
hardware driver could
not be instantiated.
The PTP driver failed on initialization.
–310 System error; The PTP
hardware driver
reported a
configuration error.
Failure in the execution of the PTP driver. The most likely cause of this error
is a mismatch between versions of the PTP driver and the LXI middleware.
–310 System error; The PTP
ordinary clock
reported a
configuration error.
Failure in execution of the DMC libraries’ PTP controller.
–310 System error; The
Trigger alarm delayed
LAN event was not
scheduled due to an
existing pending
event.
Delayed LAN events cannot occur too close together (within 20 ms).
–310 System error; The
Trigger alarm delayed
LAN event was not
scheduled due to a
conflict with an
existing scheduled
alarm.
Delayed LAN events cannot occur too close to a scheduled Alarm (within
20 ms).
–310 System error; The
Trigger alarm was not
scheduled due to a
conflict with an
existing scheduled
alarm.
Alarms cannot be scheduled to happen too close together (within 20 ms).
–311 Memory error There is a physical problem with the instrument memory, such as a parity
error.
Err# Message Verbose/Correction Information

Table of Contents

Other manuals for Keysight Technologies N9030B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N9030B and is the answer not in the manual?

Keysight Technologies N9030B Specifications

General IconGeneral
BrandKeysight Technologies
ModelN9030B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals